Welch Allyn Blood Pressure Machine Error Codes

Contents

Message: Replace RRa probe. Communications Module Messages: Message: Communications module did not power on properly. Apply a new probe cover. 3. Error Code: E7 - Internal SpO2 error. http://txtbl.com/welch-allyn/welch-allyn-blood-pressure-error-codes.html

Action: Call for service to modify advanced settings. If you continue to get errors call Medicanix 203-324-3711 for technical help. Temperature Operation 1. Message: Insert correct color-coded probe well.

Welch Allyn Error Codes

Contact Technical Service. C03 Inflation too rapid: Check for kinked hose or air obstruction. Air leak. Retake reading.

Skip navigation UploadSign inSearch Loading... SpO2 Operation 1. If problem persists contact Technical Service. Welch Allyn Spot Vital Signs Monitor Action: Power down and restart.

Oldest record overwritten. Action: Check that the calibration gas has the correct concentration. Contact Technical Service. https://www.manualslib.com/manual/826624/Welch-Allyn-Spot-Vital-Signs-Lxi.html?page=47 Spot LXI Error Code List: Error Code: C12 - Device outside operating temperature range.

Check for hose constriction. Welch Allyn Spot Vital Signs Lxi Set Date And Time Action: Check the connections and tubing for kinks. C03 Inflation too rapid: Check for kinked hose or air obstruction. Apply sensor to patient and verify pulse signal bar graph. 2.

Welch Allyn Spot Vital Signs 420 Series

Contact Technical Service.

E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.C22 10-minute diagnostic limit exceeded: Check patient. Welch Allyn Error Codes Action: Reattach the cable to the monitor. Welch Allyn Spot Vital Signs Battery B Error Codes When responding to an error code, always check the patient first and then the device.

Error Code: Av! - Device outside operating temperature range. http://txtbl.com/welch-allyn/welch-allyn-error-codes-c05.html Mode Button 1. Action: Power down and restart. C03 Inflation too fast Your BP hose going to the cuff may be kinked causing an air obstruction. Welch Allyn Spot Vital Signs E42

If problem persists, return for servicing.* Push Start/Stop button to retry reading. Causes: The NIBP module has an air leak. Change ambient temperature. this content Read the Directions for Use.

Action: Check the sampling line for kinks or obstructions. Welch Allyn Error Code E43 Press again to cancel the blood pressure measurement. Action: Resume using the monitor.

C8 Bad sensor: Replace sensor.

Action: Check connections; limit patient movement. To delete all unsaved data, touch Clear. Read the Directions for Use. Welch Allyn C05 Error Message: Unable to determine NIBP; check inflation settings.

Replace probe. Error Code: C03 - Inflation too rapid. Internal malfunction. http://txtbl.com/welch-allyn/welch-allyn-blood-pressure-monitor-error-codes.html Internal malfunction.

Message: External device not recognized. Causes: There is excessive patient movement (Low confidence condition) Action: Check the patient. Check all tubing connections for leaks. Action: Add a patient ID.

Causes: The battery is faulty. Press to start a blood pressure measurement. 2. SpO 2 Error Codes C6 SpO 2 heart rate out of range. Temperature Error Codes C20 Broken/missing probe: Replace probe.

Check patient, contact Technical Service Error Code: C12 - Ambient temperature out of range. C8 Bad sensor: Replace sensor. Press again to scroll through presets. Causes: The ambient temperature is out of range.

C12 Ambient temperature out of range: Adjust room temperature or device location. Welch Allyn Spot Vital Signs Monitor Error Code Guide   Welch Allyn Spot Vital Signs Monitor Error Code Guide You might find the following table helpful for troubleshooting error codes After one blood pressure measurement the Spot Vital Signs returns to the default inflation pressure. Message: Retry temperature measurement. (Note: This message often accompanies other temperature messages.) Causes: A probe heater or data error occurred.

Action: Touch OK to dismiss or connect the monitor to AC power. Set SpO2 alarm limits to OFF. C04 Excessive inflation time: Check for air leaks. Action: Replace the temperature probe.

SpO 2 Error Codes C6 SpO 2 heart rate out of range.